Setup
3.1 Sensor settings
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
33
Setup → Sensor settings → User calibration → Path 1 offset
Multipoint calibration
3.1.9.1
Enable multipoint calibration
Description
Enable or disable the multipoint calibration.
Setup → Sensor settings → Multipoint calibration
Asymmetric calibration table
Description
Note
Access level: Expert user only
Only available for LUILOCAL_CURRENT_LUI_ACCESS_LEVEL == SUP AND
PID_ProtectCalibrationTable == 0 AND PID_EnableUserCalibrationTable == 1
Usage of the calibration table:
● Unidirectional: All 20 calibration points are used for positive flow calibration and are then
mirrored for negative flow.
● Bidirectional: The first 10 calibration points are used for positive flow calibration and the
10 last calibration points are used for negative flow calibration.
Unidirectional; Bidirectional
Setup → Sensor settings → Asymmetric calibration table