Setup
3.1 Sensor settings
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
33
Related
Path
Setup Sensor settings User calibration Path 1 offset
3.1.9
Multipoint calibration
3.1.9.1
Enable multipoint calibration
Description
Enable or disable the multipoint calibration.
Setting
Disabled; Enabled
Default
Disabled
Related
Path
Setup Sensor settings Multipoint calibration
3.1.9.2
Asymmetric calibration table
Description
Note
Access level: Expert user only
Only available for LUILOCAL_CURRENT_LUI_ACCESS_LEVEL == SUP AND
PID_ProtectCalibrationTable == 0 AND PID_EnableUserCalibrationTable == 1
Usage of the calibration table:
Unidirectional: All 20 calibration points are used for positive flow calibration and are then
mirrored for negative flow.
Bidirectional: The first 10 calibration points are used for positive flow calibration and the
10 last calibration points are used for negative flow calibration.
Setting
Unidirectional; Bidirectional
Default
Unidirectional
Related
Path
Setup Sensor settings Asymmetric calibration table