Setup
3.4 Inputs and outputs
FST020 Function manual
78 Function Manual, 09/2017, A5E42949526-AA
Setting
195 CH2 loop cur. in lower saturation;
196 CH2 loop cur. in upper saturation;
197 CH2 cable break;
209 CH4 cable break;
210 CH4 output frequency too low;
211 CH4 output frequency too high;
212 CH4 pulse overflow;
214 Channel 2 simulated;
216 Channel 4 simulated;
217 Process values frozen;
218 All outputs forced;
219 CH2 loop current deviation;
222 Invalid register mapping;
223 Invalid coil configuration
Default
None
Related
Path
Setup Inputs and outputs Channel 3 - relay Input/output alarms
3.4.3.11
Simulation alarms (1)
Description
Selection of alarm items that impact the output. Mainly simulation alarms. Applicable if
Status mode = Alarm item.
Setting
160 Mass flow simulated; 161 Volume flow simulated; 162 Density simulated; 163 Medium
temp. simulated; 167 Totalizer 1 simulated
Default
None
Related
Path
Setup Inputs and outputs Channel 3 - relay Simulation alarms (1)
3.4.3.12
Simulation alarms (2)
Description
Selection of alarm items that impact the output. Applicable if Status mode = Alarm item.
Setting
289 Sound velocity simulated; 290 Flow velocity simulated; 292 Kin. viscosity simulated;
300 Std. Kin. viscosity simulated
Default None
Related
Path Setup Inputs and outputs Channel 3 - relay Simulation alarms (2)