Setup
3.1 Sensor settings
FST020 Function manual
26 Function Manual, 09/2017, A5E42949526-AA
3.1.6.2
Clamp-on sensor size
Description
Installed sensor size. Read only. Can only be changed via the Sensor Wizard. Indicates the
Sensor model chosen from the sensor setup wizard.
3.1.6.3
Nominal sensor frequency
Description
Nominal frequencies for the sensors
Setting 80000.0 m/s to 4000000.0 m/s
Default
1100000.0 m/s
Related
Path
Setup Sensor settings Sensor selection Nominal sensor frequency
3.1.6.4
Sensor crystal projection
Description
Length of the crystal projection onto the emitting surface of the sensor (clamp-on). Access
level: Expert user only.
Setting
0.0 m to 0.3 m
Default 0.03409 m
Related
Path
Setup Sensor settings Sensor settings Sensor crystal projection
3.1.6.5
Sensor phase velocity
Description
The velocity at which the wave fronts traverse the interface between the wedge and pipe.
Access level: Expert user only.
Setting
1000.0 m/s to 20000.0 m/s
Default
3695.5 m/s
Related
Path
Setup Sensor settings Sensor settings Sensor phase velocity