Setup
3.1 Sensor settings
FST020 Function manual
30 Function Manual, 09/2017, A5E42949526-AA
Related
Path
Setup Sensor settings Path settings Path 1 Path geometry
3.1.7.2.3 Sensor spacing
Description
Sensor spacing - Distance between the front faces of the sensors for path 1.
Setting -100.0 m to 100.0 m
Default
0.0 m
Related
Path
Setup Sensor settings Path settings Path 1 Sensor spacing
3.1.7.2.4 Zero point adjustment
Description
Zero point adjustment. Can only be changed via the Sensor Wizard.
Note
Important
Please stop all flow in the pipe and perform this function. The flow velocity is measured prior
and after the function has been applied.
Setting
5 to 300
Default
10
Related
Path
Setup Sensor settings Path settings Path 1 Zero point adjustment Time duration
3.1.7.2.5 Serial number
Description
Serial number of sensor (clamp-on)
Setting
Default