Setup
3.4 Inputs and outputs
FST020 Function manual
68 Function Manual, 09/2017, A5E42949526-AA
Example
Pulse output configuration (channels 2 to 4)
Operation mode = Pulse output
Process value = Mass flow
Amount per pulse = 1 kg
Pulse width = 1 ms
Measured mass flow value = 10 kg/s (constant)
Result:
Pulse repetition = 100 ms
Output frequency = 10 pulses per second with a pulse width of 1 ms
Remaining time between pulses is 99 ms
Note
Pulse width must be selected with the view that
remaining time is always greater than pulse
width at the highest measured flow.
Frequency output
The frequency output function supplies a frequency (50% duty cycle) proportional to the
selected process value.
Frequency is calculated as follows:
Example
This example shows how to calculate the output frequency for any measured flowrate:
Frequency output configuration:
Operation mode = Frequency output (Channel 2 to 4)
Process value = Mass flow
Direction = Positive
Frequency value high = 12 kHz
Frequency value low = 2 kHz
Upper range value = 15 kg/s
Lower range value = 5 kg/s