Setup
3.4 Inputs and outputs
FST020 Function manual
82 Function Manual, 09/2017, A5E42949526-AA
3.4.3.3.3 Frequency value high
Description
Upper frequency value. Upper nominal output range boundary.
Setting
0.0 Hz to 12500.0 Hz
Default
10000.0 Hz
Related
Path
Setup Inputs and outputs Channel 4 - input/output Frequency output Frequency
value high
3.4.3.3.4 Frequency value low
Description
Lower frequency value. Lower nominal output range boundary.
Setting
0.0 Hz to 12500.0 Hz
Default
0.0 Hz
Related
Path
Setup Inputs and outputs Channel 4 - input/output Frequency output Frequency
value low
3.4.3.3.5 Upper range value
Description
Sets value at 100% of measuring span and corresponds to Frequency value high.
Setting and Default are sensor specific, depending on the setting in Process value and
depending on the Units setting.
Setting
-10800000000.00 kg/h to +10800000000.00 kg/h
Default
+10800000000.00 kg/h
Related
Path
Setup Inputs and outputs Channel 4 - input/output Frequency output Upper range
value
3.4.3.3.6 Lower range value
Description
Sets value at 0% of measuring span and corresponds to Frequency value low.