SVAN 948 USER MANUAL
4 - 77
The view of the displays in the BUFFERS SETUP sub-list; the all the parameters were selected
to be saved in the buffer’s file; in CHANNEL 1 and 2 vibration and in CHANNEL 3 and 4 sound parameters
Notice: The change of the channels parameters is not possible when the
measurement is performed. The user has to finish the current measurement.
Notice: Setting BUFFER:OFF results in bypassing the storing of the given channel’s results
in the buffer’s file. It creates more space in the buffer for other channels (see Chapter 5 for details).
Notice: The buffer is updated in the time periods (steps) defined in the BUF. STEP (path:
MENU / INPUT / MEASURE SETUP / BUF. STEP). The way the measurement results are stored is described
in Chapter 5.
Selection of 1/3 OCTAVE analysis parameters for vibration signal - SPECTRUM
The SPECTRUM sub-list (path: MENU/ INPUT/ 1/3 OCTAVE SETUP/ CHANNEL x/ SPECTRUM) is a
context element, which appears on the INPUT sub-list in 1/1 OCTAVE or 1/3 OCTAVE mode.
The view of the displays in the 1/3 OCTAVE SETUP window with the SPECTRUM text
selected in CHANNEL x position
User can switch OFF or ON SPECTRUM position in selected channels by pressing the <
> or
<
> push-buttons. After pressing the <ENTER> push-button any changes made in the sub-list are
confirmed and the sub-list is closed. The return to the INPUT sub-list ignoring any changes made in the
sub-list is made after pressing the <ESC> push-button
Notice: CHANNEL 2, CHANNEL 3 and CHANNEL 3 position are not available. The user
can not ON this CHANNEL to display SPECTRUM.