SVAN 948 USER MANUAL .
3 - 14
v SETUP (one of the main lists available after pressing the <MENU> push-button)
TIMER (sub-list)
§ TIME (position); it enables the user to set time of the self switching on of the instrument
RTC (sub-list)
§ RTC (position); it enables the user to set the internal real time clock of the instrument
FIELD CORRECTION (sub-list available only in the case of sound measurements)
§ FIELD CORRECTION (position); available values of the compensation required in the case of
the measurements performed in the diffuse field conditions: FREE or DIFFUSE
USER FILTERS (sub-list)
§ VIBR. FILTERS
• VIEW
o VIEW enables the user to select which filter should be viewed; the available
options are VUSR1, VUSR2, VUSR3 and any other transmitted to the instrument
from a PC by means of the interface
After pressing the <ENTER> push-button another sub-list opens containing the
values of filters used in 1/1 OCTAVE or 1/3 OCTAVE analysis and saved under
the name USER1, USER2, USER3 or any other
• EDIT
o EDIT enables the user to select which filters should be edited; the available
options are as follows: VUSR1, VUSR2, VUSR3 or any other transmitted to the
instrument from a PC by means of the interface
o After pressing the <ENTER> push-button another sub-list opens containing the
values of the filters used in 1/1 OCTAVE or 1/3 OCTAVE analysis; the user can
set the values of correcting coefficients for all 1/3 OCTAVE filters:
0.80 Hz: available values of 0.8 Hz centre frequency filter: -100.0dB ...
100.0dB
...
…
20.0kHz: available values of 20 kHz centre frequency filter: -100.0dB ...
100.0dB
• CLEAR (position)
o CLEAR enables the user to select which filters should be cleared; the available
options are as follows: ALL, VUSR1, VUSR2, VUSR3 or any other
§ SOUND FILTERS
• VIEW
o VIEW enables the user to select which filter should be viewed; the available
options are SUSR1, SUSR2, SUSR3 and any other transmitted to the instrument
from a PC by means of the interface
After pressing the <ENTER> push-button another sub-list opens containing the
values of filters used in 1/1 OCTAVE or 1/3 OCTAVE analysis and saved under
the name USER1, USER2, USER3 or any other
• EDIT
o EDIT enables the user to select which filters should be edited; the available
options are as follows: SUSR1, SUSR2, SUSR3 or any other transmitted to the
instrument from a PC by means of the interface
o After pressing the <ENTER> push-button another sub-list opens containing the
values of the filters used in 1/1 OCTAVE or 1/3 OCTAVE analysis; the user can
set the values of correcting coefficients for all 1/3 OCTAVE filters:
0.80 Hz: available values of 0.8 Hz centre frequency filter: -100.0dB ...
100.0dB
...