SVAN 948 USER MANUAL .
4 - 6
The lowest value of the sensitivity to be able to introduce is equal to 50.0 µV / Pa (it conforms to the
calibration factor equal to 60.0 dB) and the greatest one - 50.0 V / Pa (calibration factor equal to -60.0 dB).
In order to return to the CALIBRATION sub-list the user has to press the <ESC> push-button.
a)
b)
The view of the displays with the lowest possible sensitivity and the greatest calibration factor (a)
and the greatest sensitivity and the lowest calibration factor (b)
Notice: The calibration factor is always added to the results of sound level measurements
and sound analysis.