8.1.8 Connection modes of test terminals
TH2840X adopts 4-terminal test method.
Hcur: current sample high terminal
Lcur: Current sample low terminal
Hpot:voltage sample high terminal
Lpot: Voltage sample low terminal
8.1.9 Test speed (Frequency>=10kHz)
Fast+:about 1800 times/s(0.55ms/time)
Fast:about 300 times/s(3.3ms/time)
Medium:about 11 times/s(90ms/time)
Slow:about 4times/s(240ms/time)
The fast+, fast and middle speed will be slow down when frequency <10kHz.
8.1.10 Average
255 programmable
8.1.11 Display digit
6 digits, max. displayed digit: 999999
8.2 Test signal
8.2.1 Test signal frequency
Test signal is sine wave, accuracy:0.01%
Frequency range:
20Hz~500kHz(TH2840A/TH2840AX/TH2840NX)
20Hz~2MHz(TH2840B/TH2840BX)
Min. resolution:1mHz
8.2.2 Signal mode
Normal: When testing, on measurement display page, voltage across test terminals may be smaller
than preset voltage.
Constant level: The auto adjustment of internal level makes the voltage of DUT accordant with
preset voltage.