Tonghui TH2840 Series - Handler Subsystem Commands

Tonghui TH2840 Series
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
157
Example:
:TSSE:IMP CP,CS,LP,LS ----Set 4 parameter functions at the same time
:TSSE:IMP1 RP ----Set parameter 1 to RP
:TSSE:IMP2 RS ----Set parameter 1 to RS
:TSSE:IMP3 Z ----Set parameter 1 to Z
:TSSE:IMP4 Y ----Set parameter 1 to Y
:TSSE:IMP1? ----Query parameter 1 function
:TSSE:IMP2? ----Query parameter 2 function
:TSSE:IMP3? ----Query parameter 3 function
:TSSE:IMP4? ----Query parameter 4 function
:TSSE:IMP? ---- Query the 4-parameter function
9.2.15 Handler subsystem commands
9.2.15.1 LCR handler subsystem commands
Description: Set LCR Handler mode
Command systax:
:HAND:STAT?
:HAND:STAT <0|1|2|OFF|ON|BUS>
Parameter:
0|OFF ---- means default
1|ON ---- means custom
2|BUS ---- means bus control
Example:
:HAND:STAT 0 ----Set the HANDLE mode as default;
:HAND:STAT OFF ----Set the HANDLE mode as default;
:HAND:STAT 1 ----Set the HANDLE mode to custom;
:HAND:STAT ON ----Set the HANDLE mode to custom;
:HAND:STAT BUS ---- Return to the HANDLE mode to BUS.
Description: Set LCR Handler custom output control
Command systax:
:HAND:STAT:OFfLO <…>
:HAND:STAT:OFfHI <…>
:HAND:STAT:LVLO <…>
:HAND:STAT:LVHI <…>
:HAND:STAT:PUHI <…>
:HAND:STAT:PULO <…>
Parameter:
<…> is a list format, such as 1, 2, 4, 7, 14, 24, etc., corresponding to the table position displayed by
the instrument
Example:
:HAND:STAT:OFLO 1,3,5,9 ----Set the 1,3,5,9 index function as constant low level;

Table of Contents