0|AUTO ----AUTO
1|FIX----FIX
Example:
:BIAS:POL:AUTO 0 ----Set the bias polarity to AUTO
:BIAS:POL:AUTO 1 ----Set the bias polarity to FIX
:BIAS:POL:AUTO? ----Return to bias polarity status
9.2.8 TRIGger subsystem commands
The TRIGger subsystem commands are mainly used to set the instrument trigger source, trigger
delay and trigger measurement.
Description: Trigger the instrument to measure once
Command systax:
:TRIG
Parameter:
Example:
:TRIG ---- Trigger the instrument to measure once
9.2.8.1 Trigger mode
Description: Set the trigger source mode of the instrument
Command systax:
:TRIG:SOUR?
:TRIG:SOUR<CONT | SING>
Parameter:
:CONT ----Continuous
:SING ----Single
Example:
:TRIG:SOUR CONT ----Set to continuous trigger
:TRIG:SOUR SING ----Set to single trigger
:TRIG:SOUR? ----Return tothe trigger mode
9.2.8.2 Trigger delay
Description: Set the trigger delay time of the instrument
Command systax:
:TRIG:DEL?
:TRIG:DEL<float | MIN | MAX>
Parameter:
float ---- Represent the floating-point data
MIN ---- Set the minimum settable value
MAX ---- Set the maximum settable value
Example: