EasyManua.ls Logo

Tonghui TH2840 Series - TRACE Subsystem Commands

Tonghui TH2840 Series
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
154
f
n
---- The floating point number of the sweep point n
f
n+1
---- The floating point number of the sweep point n+1
f
n+2
---- The floating point number of the sweep point n+2
f... ---- The floating point number of the sweep point n…
Example:
:LIST:DEL 0.01, 20m, 30m ---- Set the test delay of the first 3 points;
:LIST:DEL 1m, 2m, 3m, 4m, 2m ---- Set the test delay of the first 5 points;
:LIST:DEL6 1m, 2m, 3m, 4m, 5m ---- Set the test delay of 6~10 points;
:LIST:DEL25 20m ---- Set the test delay of the 25th point;
:LIST:DEL? ----Return to the test delay of all sweep points;
:LIST:DEL4? ----Return to the test delay of point 4.
9.2.14 TRACE subsystem commands
9.2.14.1 Sweep Points
Description: Set the number of LCR curve scan points
Command systax:
:TSSE:POINT?
:TSSE:POINT <51|101|201|401|801>
Parameter:
51 ---- means scanning 51 points
101 ---- means scanning 101 points
201 ---- means scanning 201 points
401 ---- means scanning 401 points
801 ---- means scanning 801 points
Example:
:TSSE:POINT 51 ----Set LCR to scan 51 points
:TSSE:POINT 101 ----Set LCR to scan 101 points
:TSSE:POINT 201 ----Set LCR to scan 201 points
:TSSE:POINT 401 ----Set LCR to scan 401 points
:TSSE:POINT 801 ----Set LCR to scan 801 points
:TSSE:POINT? ----Query the LCR sweep points
9.2.14.2 Sweep Type
Description: Set the sweep parameter type of LCR curve
Command systax:
:TSSE:MODE?
:TSSE:MODE <FREQ|VOLT|CURR|BVOL|BCUR>
Parameter:
Enumeration type string parameter;
Example:

Table of Contents

Related product manuals