3.5.6 BIN Switch
Parameter setting attributes: enumerated type.
Set the independent comparison switch of the specific sorting BIN:
Turn off the comparison function of the specified BIN
Turn on the comparison function of the specified BIN
When the corresponding sorting BIN is OFF, the sorting process will skip this sorting limit
comparison.
3.5.7 High-low limit
As the main basis for parameter comparison.
If the upper and lower limits of the corresponding BIN are not set, it indicates that the parameter in
the corresponding BIN does not participate in the comparison, that is, the test result does not affect
the comparison result; if only a lower limit or upper limit is set, the effect of unilateral comparison
will be produced.
Parameter setting attribute: numeric input type.
Note: When the upper and lower limits only set the upper or lower limit, it is regarded as a unilateral
comparison.
3.6 <List Setup>
Press [SETUP] and then List Setup to enter into the <List Setup> page as shown below.
(Figure 3-6-1 List setup)
The list sweep function can perform auto sweep test for the test frequency, test level or bias voltage
of 201 points. On <LIST SETUP> page, the following list sweep parameters can be set.
Total Point
Trigger mode
List mode (sweep mode)