145
reference value
:FUNC:DEV4:REF? ----Return to the deviation reference value of parameter 4
:FUNC:DEV:REF? ----Return to the deviation reference values of 4 parameters
:FUNC:DEV:REF 10,11,12,13 ----Set the deviation reference values of the 4-parameters at the same
time
9.2.11.6 Step Delay
Description: Set the instrument step delay time
Command systax:
:FUNC:SDEL?
:FUNC:SDEL<float | MIN | MAX>
Parameter:
float ---- Represents the floating-point data
MIN ---- Set the minimum settable value
MAX ---- Set the maximum settable value
Example:
:FUNC:SDEL 0 ----Set the step delay to 0;
:FUNC:SDEL 1ms ----Set the step delay to 1ms;
:FUNC:SDEL 1 ----Set the step delay to 1s;
:FUNC:SDEL? ----Return the step delay value;
9.2.12 COMParator subsystem commands
The COMParator subsystem commands are used to set the bin comparator function, including the
setting of the comparison switch and the setting of the limit list.
9.2.12.1 COMP comparison switch
Description: Set the instrument comparison function on or off
Command systax:
:COMP? :COMP<0|1|ON|OFF >
Parameter:
0|OFF ---- OFF
1|ON ---- ON
Example:
:COMP 0 ---- Set the comparison function to OFF
:COMP 1 ---- Set the comparison function to ON
:COMP? ----Return to compare function switch status
9.2.12.2 COUNt count switch
Description: Set the instrument comparison counting function ON or OFF