9.2.12.6 CLRar Clear Table
Description: Used to clear the limit setting data of each bin.
Command systax:
:COMP:BIN:CLE
Parameter:
Example:
:COMP:BIN:CLE ----Clear the limit data in the table
9.2.12.7 BIN:SW Bin switch
Description: Set the specified bin comparison function to on or off.
Command systax:
:COMP:BIN<n>:SW?
:COMP:BIN:SW<0|1|ON|OFF >
Parameter:
n ---- bin number index, the value is 1~10
0|OFF ---- OFF
1|ON ---- ON
Example:
:COMP:BIN1:SW 0 ---- Set the comparison function of bin 1 to OFF
:COMP:BIN2:SW 1 ----Set the comparison function of bin 2 to ON
:COMP:BIN10:SW? ----Return to BIN10 comparison function switch status.
9.2.13 LIST subsystem commands
The LIST subsystem commands are mainly used to set the list sweep function, sweep points, sweep
mode, sweep limits.
9.2.13.1 TOTAL sweep points
Description: Set the total number of scan points in the list
Command systax:
:LIST:TOTAL?
:LIST:TOTAL <int>
Parameter:
int ----The total number of scan points, the value is 1~201
Example:
:LIST:TOTAL 8 ----Set the list sweep points to 8
:LIST:TOTAL 201 ----Set the list sweep points to 201
:LIST:TOTAL? ---- Return to the number of sweep points in the list