155
:TSSE:MODE FREQ ----Set the LCR sweep parameter to frequency
:TSSE:MODE VOLT ----Set the LCR sweep parameter to voltage
:TSSE:MODE CURR ----Set the LCR sweep parameter to current
:TSSE:MODE BVOL ----Set LCR sweep parameter as bias voltage
:TSSE:MODE BCUR ----Set LCR sweep parameter as bias current
:TSSE:MODE? ----Query the LCR sweep parameter type.
9.2.14.3 Sweep Range
Description: Set the LCR sweep range (start and stop points)
Command systax:
:TSSE:SWEEP?
:TSSE:SWEEP <float,float>
Parameter:
float ---- float string format, such as "3.14159";
Example:
:TSSE:SWEEP 1,100 ----Set the LCR sweep start and stop point
:TSSE:SWEEP? ----Query the LCR sweep start and stop point
9.2.14.4 Sweep Mode
Description: Set the LCR sweep mode
Command systax:
:TSSE:SMODE?
:TSSE:SMODE <SEQ|STEP>
Parameter:
SEQ ---- continuous mode
STEP ----single step mode
Example:
:TSSE:SMODE SEQ ----Set the LCR sweep mode to continuous
:TSSE:SMODE STEP ----Set the LCR sweep mode to single step
:TSSE:SWEEP? ----Query the LCR sweep mode
9.2.14.5 Coordinate Mode
Description: Set the LCR curve coordinate mode
Command systax:
:TSSE:FORMat?
:TSSE:FORMat <LIN|LOGX>
Parameter:
LIN ---- Linear coordinates
LOGX ---- X-axis logarithm