:LIST:FUNC:IMPB5 CP,CS ----Set the parameter function of point 5~6 of the second parameter
:LIST:FUNC:IMPC? ----Return to all point parameter functions of the third parameter
:LIST:FUNC:IMPD11? ---- Return to the parameter function of the 11th point of the fourth
parameter.
Description: Linear setting of the specified parameter
Command systax:
:LIST:FUNC:LIN<A|B|C|D><start,stop,para>
Parameter:
A|B|C|D ---- Indicate that one of the 4 parameters is specified
start ---- Specify the starting point of the setting, the value is 1~201
stop --- Specify the end point of the setting, the value is 1~201
para ---- function name (refer to the table above)
Example:
:LIST:FUNC:LINA 1,15,LP ----Set the function of point 1~15 of the first parameter to LP
:LIST:FUNC:LINB 5,201,CP ----Set the function of point 5~201 of the second parameter to CP
:LIST:FUNC:LIND 3,15,LP ----Set the function of point 3~15 of the fourth parameter to LP
9.2.13.10 BAND Limit
Description: Set the limit data in the specified point in list sweep setting table
Command systax:
:LIST:BAND<n>?
:LIST:BAND<n> OFF
:LIST:BAND<n><A|B|C|D,lo,hi>
Parameter:
n ---- Specify the index of the sweep point, ranging from 1 to 201
A|B|C|D ---- Indicate that one of the 4 parameters is specified
lo ---- the lower limit of the parameter corresponding to the specified point
hi ---- the upper limit of the parameter corresponding to the specified point
Example:
:LIST:BAND1 OFF ----Clear the limit data of the 4parameters of point 1
:LIST:BAND2 A,1,2 ----Set the function of point 5~201 of the second parameter as CP
:LIST:BAND201 D,1.1,2.2 ----Set the function of point 3~15of the 4th parameter as LP
:LIST:BAND9? ---- Return to the upper and lower limits of the 4 parameters of point 9, loA, hiA…
9.2.13.11 Delay
Description: Set the test delay of the sweep point, set the delay time of several points starting
from the nth point
Command systax:
:LIST:DEL[n]?
:LIST:DEL[n] <fn>[,fn+1][,fn+2]…
Parameter:
n ---- Start from the nth point