131
Speed Slow
Short correction should be made before testing. Connect standard DC resistors: 0.1Ω, 1Ω, 10Ω,
100Ω, 1kΩ, 10kΩ, 100kΩ. The error between the instrument reading and the standard value should
be within the allowable error range specified in this chapter for DCR accuracy.