164
:CORR:DATE DCR ---- Query the time of the last DCR correction operation
:CORR:DATE DCROPEN ---- Query the time of DCR open correction operation
:CORR:DATE DCRSHORT ---- Query the time of the DCR short correction operation
:CORR:DATE1 LAST ---- Query the time of the last point frequency 1 correction operation
:CORR:DATE2 OPEN ---- Query the time of point frequency 2 open correction operation
:CORR:DATE3 SHORT ---- Query the time of point frequency 3 short correction operation
:CORR:DATE10 LOAD ---- Query the time of point frequency 10 load correction operation
Note: The format of the returned data is "yyyy-MM-dd hh:mm:ss", and a return of "----" indicates
that the date is invalid, that is, the correction operation has not been performed.
9.2.17.8 SPOT<n>
Description: Set the switch state of the specified frequency point
Command systax:
:CORR:SPOT<n>:STAT?
:CORR:SPOT<n>:STAT <0|1|ON|OFF>
Parameter:
n ---- subscript of frequency point index, value 1~10
0|OFF ---- OFF
1|ON ---- ON
Example:
:CORR:SPOT1:STAT 0 ----Set the frequency point 1 status to OFF
:CORR:SPOT3:STAT 1 ----Set the frequency point 3 state to OFF
:CORR:SPOT10:STAT? ----Return the status switch of frequency point 10
Description: used to specify the measurement frequency of the frequency point
Command systax:
:CORR:SPOT<n>:FREQ?
:CORR:SPOT<n>:FREQ <float | MIN | MAX>
Parameter:
float ---- Represent the floating-point data
Example:
:CORR:SPOT1:FREQ 1200 ----Set the frequency to 1200Hz;
:CORR:SPOT2:FREQ 1100HZ ----Set the frequency to 1100Hz;
:CORR:SPOT3:FREQ 1.2K ---- set the frequency to 1200Hz;
:CORR:SPOT4:FREQ? ---- Return to the frequency of frequency point 4
Description: Perform open-circuit correction for instrument specific frequency points
(frequency 1, frequency 2...)
Command systax:
:CORR:SPOT<n>:OPEN
Parameter:
Example:
:CORR:SPOT1:OPEN ---- Perform open circuit correction on frequency point 1
:CORR:SPOT2:OPEN ---- Perform open circuit correction on frequency point 2