43
Short
Perform a short correction test on the current frequency.
Load
Perform a load correction test on the current frequency.
3.8.6.1 Steps of load correction
1. Move the cursor to the Freq setting area, and set the frequency to be calibrated.
2. Open the test fixture, and press the soft key Open to perform open correction for the current set
frequency.
3. Short-circuit the test fixture and press the soft key Short to perform short-circuit correction for
the current set frequency.
4. Prepare a measurement standard device, move the cursor to the Load Type setting field, and set
the type of function parameters that the standard device needs to measure.
5. Move the cursor to the corresponding frequency setting domain, connect the standard device to
the test fixture, press the softkey Load, and the instrument will perform a load calibration. The
actual test result of the standard device is displayed on the load.