User Guide Ver. 7.1.0 2100 Cap Inspector Creating a Strip Torque Test • 59
Strip Slow Down At
Observe the minimum and maximum
values.
To input a value, just start typing. The first key
selected will overwrite the first highlighted
value on the screen. When the desired value
has been keyed in press OK to save the new
value.
Strip Pass / Fail Limit
Low Limit: The Torque value that must be
exceeded for the test to pass.
High Light: The torque value that must not be
exceeded for the test to pass.
Report Options
The Header and Comment will be displayed on all test
reports for this test profile.