96 • Cap Inspector Calibration User Guide Ver. 7.1.0 2100 Cap Inspector
Cap Inspector Calibration
What's Inside
This chapter describes the entire calibration process, from
the initial verification through the performance of a user
calibration.
General Calibration Information
A factory Calibration is burned into read-only memory when
a system is shipped or serviced by a Vibrac trained
individual. This insures the user that there is a full back up
available if a faulty user calibration is performed.
Calibration Frequency
When a system is used on a daily basis, the calibration
should be verified at least twice a year.
Calibration Equipment
In accordance with ASTM Standard D3474, a round disk, a
pulley supporting fixture and a series of weights are required
to calibrate torque meters used in the packaging industry.