2-24
IM 760201-01E
Continuous Integration Mode
Integration continues for a specified amount of time, and then resets and starts again.
Integration repeats until you press STOP. If the integrated value reaches its maximum or
minimum displayable value before the specified amount of time has elapsed, integration
is stopped, and the integration time and integrated value at that point are held.
Start
Reset
Integration time
Integrated value
Hold
Hold
Timer
value
Timer
value
Timer
value
Press
STOP
Real-Time Normal Integration Mode
You specify the dates and times when integration will start and end. When the specified
stop time is reached, or when the integrated value reaches its maximum or minimum
displayable value, integration is stopped, and the integration time and integrated value at
that point are held.
Integration time
Integrated value
Hold
Hold
Start
date/time
End
date/time
Reset
2.6 Integration