2-25
IM 760201-01E
Features
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Real-Time Continuous Integration Mode (Continuous integration)
You specify the dates and times when integration will start and end, and integration
repeats within the start and end times at the interval specified by the timer. When the
amount of time specified by the timer elapses, integration resets and starts again. When
the specified stop time is reached, or when the integrated value reaches its maximum
or minimum displayable value, integration is stopped, and the integration time and
integrated value at that point are held.
Integration time
Integrated value
Hold
Hold
Timer
value
Timer
value
Timer
value
Start
date/time
End
date/time
Reset
2.6 Integration