B. Enter your Operator ID and sequence through the workow to view the Firmware Delivery opon.
Verify Enabled is selected.
5. The baery charge level is no less than 51%.
To verify:
A. From the Home Screen, navigate to: More Opons > Instrument Status
B. Locate Baery and verify level is 51% or higher.
Procedure:
Perform the following steps to download and install soware update:
CAUTION: During the sow
are installaon, do not remove the baery from the instrument. When the
installaon is complete, the instrument will display a message indicang that new soware has been
installed. The display of this message is the indicaon that it is safe to resume normal operaon of the
instrument.
1. From the Home screen, navigate to: More Opons > Instrument Opons > Soware Installaon >
Check for Update
2. Aer the soware download is completed, installaon of soware can begin. Please skip step 3.
3. Soware download and installaon can be performed at dierent mes if needed. The iSTAT Alinity
will display the following symbol
on the Home screen in the lower right hand corner if
sow
are is downloaded but not installed. Touch this icon to display the Soware Installaon menu.
Touch the Install Pending buon to install the soware.
4. Follow the on-screen instrucons to connue with soware installaon.
5. When presented with the name of the soware package to be installed, record the name before
proceeding (this informaon may be used at the end of the procedure to verify that the soware
update has been completed).
Press Next to connue with the installaon of the soware package, OR
Press Cancel to terminate the installaon
6. Connue to follow the on-screen instrucons. During the installaon process, the instrument screen
will go blank and remain blank for several minutes. This is normal.
Do not remove the instrument from the Base Staon
7. When the installaon is complete, the instrument will display a message indicang that new soware
has been installed. The display of this message is the indicaon that it is safe to resume normal
operaon of the instrument.
106
i-STAT Alinity — System Operaons Manual Art: 754496-01 Rev. E Rev. Date: 02-Nov-2022