Item Prompt on Analyzer Required. Specify a prompt that is from 1 to 20 characters long and:
• does not begin with the characters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon.
Descripon Oponal.
Entry Prompt Select one of the following:
Entry, oponal
Pr
ompt the operator for an oponal entry.
Entry, mandatory with override
Prompt the operator for a mandatory entry. If the override feature
is chosen by the operator no further acon is required and the
instrument will advance to the next screen.
Entry, mandatory
Prompt the operator for a mandatory entry. Cartridge tesng
cannot connue unl informaon is entered.
The elds on this scr
een are:
Selecon List A list of oxygen delivery systems idened with a page name. Each
delivery system will have No Items, Mode or STATNotes Page
applied.
Selecon Lis
t Entry Name of the page for the delivery system. Example: BNC. Select
STATNotes Items.
No Page The delivery system does not have sengs or measurements.
Example: Room Air. Once No Page is selected, click Save.
Mode The delivery system has various modes. Example: Venlator. Once
Mode is selected, click Save.
223
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022