Introduction to the Control Module 9
1100 Series FD Reference Manual 321
Firmware Update
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the FLD. Use the F5 key
(FW Update) to enter the Update section. If you want to update the resident
firmware (together with specific main firmware revisions), select the a file
from the PCMCIA card (RESnnnn.DLB) and press execute. If you want to
update the main firmware, press F7 key (Transfer) to turn the module into
the resident mode (LED on module should blink yellow).
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the FLD. Use the F5 key
(FW Update) to enter the Update section. Select the a file from the PCMCIA
card (1321nnnn.DLB) and press execute. When the update has finished the
update, press F7 key (Transfer) to return the module into the normal mode
(LED on module should stays yellow).

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals