112 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
Wavelength Calibration
1 From the user interface start the FLD Wavelength Calibration (see
Figure 41).
Agilent ChemStation: Diagnosis - Maintenance - FLD Calibration
Control Module: System - Tests - FLD - Calibrate.
2 If a deviation is displayed, press Adjust and OK. The history table will be
updated.
NOTE
If the wavelength calibration process fails, refer to “Wavelength Calibration Failed" on
page 97.
NOTE
To look at the history table (ChemStation) start a wavelength calibration and abort
immediately. No changes are made to the calibration at this time.
NOTE
Rinse the flow cell with pure water at a minimum of 1.5 ml/min to get rid of the Glycogen
from the cell and the capillaries. When organic solvent is sequentially applied (without
rinsing), a blockage of capillaries may occur.