144 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Removing the Covers
When required For all repairs inside the detector
Tools re quired Screwdriver POZI 1 PT3
Parts required Depends on the work inside and the following procedures
CAUTION
There will be a risk of damaging hardware due to overheating when operating the
instrument without covers.
Preparations for this procedure:
• Turn off the detector.
• Disconnect the power cable.
• Remove the flow cell.
• Remove detector from stack and place it on the
workbench.
1 Press the release buttons and remove the front cover.
Unclip and remove the waste funnel assemly.