EasyManua.ls Logo

Agilent Technologies 1100 Series - Detector Indicators and Test Functions

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
The Wavelength Calibration Procedure
Steps
1 Preparation of the Glycogen Calibration Sample.
2 Preparation of the Flow Cell.
3 Wavelength Calibration.
4 Verification using “Check of Raman Band of Water" on page 104.
Preparation of the Glycogen Calibration Sample
1 To prepare 10 ml of the calibration solution you have to use 10 mg of the
Glycogen sample (a tolerance of
±20% is not critical).
2 Fill the prepared amount into a suitable bottle/vial.
3 Fill 10 ml of distilled water into the vial and shake.
4 Wait 5 minutes and shake again. After 10 minutes the solution is ready.
When required If application requires, or see Ta b l e 1 8 on page 101.
Tools re quired Laboratory balance
Parts required Glycogen Calibration Sample, Syringe, needle, sample filter and PEEK fitting
from the Accessory Kit

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals