34 1100 Series FD Reference Manual
2 First Steps with the Fluorescence Detector
Setting the Chromatographic Conditions
1 Set up the system with the following chromatographic conditions and wait
until the baseline gets stable.
Tabl e 4 Chromatographic Conditions
Mobile phases A = water = 35%
B = Acetonitrile = 65%
Column OSD-Hypersil column, 125 mm x 4 mm i.d. with 5
µm particles
Sample Isocratic standard sample, 1:10 diluted in
methanol
Flow rate 1.5 ml/min
Compressibility A (water) 46
Compressibility B (Acetonitrile) 115
Stroke A and B auto
Stop time 4 minutes
Injection volume 5 µl
Oven temperature (1100) 30°C
FLD Excitations/Emission Wavelength EX = 246 nm, EM = 317 nm
FLD PMT Gain PMT = 10
FLD Response time 4 seconds