First Steps with the Fluorescence Detector 2
1100 Series FD Reference Manual 39
Stray light acts in the same way as impurities in that it simulates background 
noise. In both cases, a higher noise level and therefore a higher limit of 
detection are obtained. This indicates that high sensitivity measurements 
should be done away from wavelength settings that have a high stray light 
background. 
Step 2: Optimize limits of detection and selectivity
To achieve optimum limits of detection and selectivity, analysts must find out 
about the fluorescent properties of the compounds of interest. Excitation and 
emission wavelengths can be selected for optimum limits of detection and best 
selectivity. In general, fluorescence spectra obtained with different 
instruments may show significant differences depending on the hardware and 
software used. 
The traditional approach is to extract an appropriate excitation wavelength 
from the UV spectrum that is similar to the fluorescence excitation spectrum 
(see Figure 11) and to record the emission spectrum. Then with an optimum 
emission wavelength determined, the excitation spectrum is acquired. 
Figure 11 Excitation and emission spectra of quinidine
Excitation spectrum 
with emission at 
440 nm, emission 
spectrum with 
excitation at 250 nm 
of 1 ug/ml quinidine.
Detector settings: 
step size 5 nm, PMT 
12, Response time 4 s.