Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 79
Shutdown
An external instrument has generated a shut-down signal (through CAN or
REMOTE lines).
The detector continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
• Leak detected in an external instrument with a remote connection to the
system.
• Shut-down in an external instrument with a remote connection to the
system.
• The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
✔ Fix the leak in the external instrument before restarting the pump.
✔ Check external instruments for a shut-down condition.
✔ Check the degasser for an error condition. Refer to the Reference Manual
for the Agilent 1100 Series vacuum degasser.