EasyManuals Logo

Agilent Technologies 1100 Series Reference Manual

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 97
Wavelength Calibration Failed
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message “Wavelength Calibration Failed” is displayed and the instrument
stays in a Not Ready condition.
Probable Causes
Flash lamp not ignited or position not correct.
Cell position not correct.
Solvent in the cell not clean or air bubble in the cell.
monochromator assembly position not correct (after replacement).
Suggested Actions
Check the flash lamp image and position. See also “No Peaks" on page 100.
Check the cell position.
Flush the flow cell.
Reset monochromator settings and re-run the wavelength calibration.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals