200 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Replacing the Detector’s Firmware
The installation of new firmware is required:
if new version solves problems of currently installed version, or
if after exchange of the main board (FLM) the version on board is older
than the previously installed one.
To upgrade the detector’s firmware the following steps have to be taken:
1 Load the firmware into the detector, see help system of your user interface.
2 If FLM board was replaced, reenter the serial number information of the
module through the user interface, see “Entering the Serial Number using
the Control Module" on page 161 or “Entering the Serial Number using the
Agilent ChemStation" on page 162.