162 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Entering the Serial Number using the Agilent ChemStation
Module serial numbers are entered by typing specific commands in the
command line at the bottom of the main user interface screen.
1 To enter a module serial number, type the following command in the
command line:
print sendmodule$(lfld,"ser YYYYYYYYYY")
Where: YYYYYYYYYY is the 10-character serial number of the module in
question.
The reply line will respond with RA 0000 SER followed by the module serial
number you just entered.
2 Turn off the detector, then on again and restart the Agilent ChemStation. If
the serial number you have just entered is different than the original
module serial number, you will be given the opportunity to edit the
configure 1100 access screen during the restart of the Agilent ChemStation.
3 After start up, the serial number you have just entered can be seen under
the Instrument menu of the main user interface screen. The serial number
of the detector can also be seen by typing the following command into the
command line:
print sendmodule$ (lfld,"ser?")
The reply line will give the module serial number.
NOTE
The first two characters are letters, which should be capitalized.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals