Optimizing the Fluorescence Detector 3
1100 Series FD Reference Manual 67
Changing the Xenon Flash Lamp Frequency
Modes
The lamp flash frequency can be changed into the following modes:
Best sensitivity can be expected with no economy”, see Figure 28.
Tabl e 16 Flash Lamp Modes
Positioning 296 Hz (Standard), 560 V 63 mJoule (18.8 W)
74 Hz (Economy), 560 V 63 mJoule (4.7 W)
Rotation (Multi Ex/Em) 74 Hz (Standard), 950 V 180 mJoule (13.3 W)
74 Hz (Economy), 560 V 63 mJoule (4.7 W)
Figure 28 Xenon Flash Lamp Frequency
Standard=296 Hz
Economy=74 Hz