138 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
How to use the Cuvette
The cuvette is used for off-line measurements (no flow system required) and is
basically a standard flow cell with a few changes:
wide bore capillary connections for easier injections with a syringe
identification lever for cell auto-recognition system.
1 Install the cuvette instead of the standard flow cell.
2 Connect the waste tubing to the outlet of the cuvette.
3 Use the syringe (see “Cuvette Kit" on page 214) to inject the compound.
4 Setup the parameters for the Fluorescence Scan (under Special Setpoints).
5 Select “Take Fluorescence Scan” on the user-interface to start the off-line
measurement.