Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 155
Replacing the Lamp Cover Window
When required If window is blocking the light throughput due to contamination
Tools re quired Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm long
Hexagonal screwdriver 4 mm, 100 mm long
Parts required Quartz Window 1000-0999
NOTE
When loosening or removing screws take care that they do not fall into the instrument. This
may result in a complete removal of other assemblies.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell or disconnect the tubings from the
flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in Removing the
Covers" on page 144.
1 Disconnect the emission and the excitation cables and
from the detector main board.
A
B
A
B
A
EM
EX