96 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
Flash Trigger Lost
This message is displayed when the flash trigger is no longer generated.
Probable Causes
firmware problem
defective encoder
Suggested Actions
Reboot the detector (power cycle).
Multi Mode Off: exchange the firmware or replace the main board.
Multi Mode Ex: replace EX monochromator assembly.
Multi Mode Em: replace EM monochromator assembly.