Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 95
Flash Lamp Current Overflow
The lamp current of the xenon flash lamp is monitored constantly. If the
current gets too high, an error is generated and the lamp is turned off.
Probable Causes
• shortage of flash lamp assembly
• shortage of trigger pack assembly.
• defective FLL board.
Suggested Actions
✔ Exchange the flash lamp assembly.
✔ Exchange the FLL board (includes the trigger pack).