Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 165
Exchanging the Leak Sensor or Leak Panel
When required If defective
Tools re quired Screwdriver POZI 1 PT3
Flat screwdriver
Parts required Leak sensor assembly 5061-3356
Leak Pan G1321-45501
Preparations for this procedure:
• Turn off the detector.
• Disconnect the power cable.
• Remove the flow cell.
• Remove detector from stack and place it on the
workbench.
• Remove the covers as described in “Removing the
Covers" on page 144.
• Remove the flow cell.
1 Locate the leak sensor cable.