176 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Replacing the Slits
When required If required.
Tools re quired Screwdriver POZI 1 PT3
A pair of tweezers
Parts required Slit EX G1321-08103
Slit EM G1321-08102 (also used for PMT)
NOTE
DO NOT touch the surface of any optical component during this procedure. This will
deteriorate the reflection process.
Preparations for this procedure:
• Turn off the detector.
• Disconnect the power cable.
• Remove the flow cell.
• Remove detector from stack and place it on the
workbench.
• Remove the covers as described in “Removing the
Covers" on page 144.
1 Carefully remove the front foam holding the foam as
shown.