Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 175
10
Replace the front foam. Next Steps:
• Replace the covers as described in “Replacing the
Foams and Covers" on page 196.
• Re-install the flow cell and the front panel
• Re-install the detector in the stack.
• Turn the lamp on.
• Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter “Wavelength Verification and Calibration" on
page 101.