78 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
Time-out
The time-out threshold was exceeded (set in sequence parameter on the
Agilent ChemStation or Configure-LC system).
Probable Causes
The analysis was completed successfully, and the time-out function
switched off the pump as requested.
A not-ready state was present during a sequence or multiple-injection run
for a period longer than the time-out threshold.
Suggested Actions
Check the logbook for the occurrence and source of a not-ready condition.
Restart the analysis where required.