14 1100 Series FD Reference Manual
1 Installing the Fluorescence Detector
Site Requirements
A suitable environment is important to ensure optimal performance of the 
detector.
Power Consideration
The detector power supply has wide ranging capabilities and accepts any line 
voltage in the range mentioned in Table 1 on page 16. Consequently, there is 
no voltage selector in the rear of the detector. There are also no externally 
accessible fuses, because automatic electronic fuses are implemented in the 
power supply.
Power Cords
Different power cords are offered as options with the detector. The female end 
of all power cords is identical. It plugs into the power-input socket at the rear 
of the detector. The male end of each power cord is different and designed to 
match the wall socket of a particular country or region.
WARNING
To disconnect the detector from line, unplug the power cord. The power supply still 
uses some power, even if the power switch on the front panel is turned off.
WARNING
Shock hazard or damage of your instrumentation can result, if the devices are 
connected to a line voltage higher than specified.
WARNING
Never operate your instrumentation from a power outlet that has no ground 
connection. Never use a power cord other than the Agilent Technologies power cord 
designed for your region.
WARNING
Never use cables other than the ones supplied by Agilent Technologies to ensure 
proper functionality and compliance with safety or EMC regulations.