172 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Replacing the Mirror Assembly
When required If reflection is too low (contaminated)
Tools re quired Screwdriver POZI 1 PT3
A pair of tweezers
Parts required Mirror assembly 1000-1000
NOTE
DO NOT touch the surface of any optical component during this procedure. This will reduce
the reflection efficiency.
NOTE
One side of the mirror is marked “R” representing the rear side.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in Removing the
Covers" on page 144.
1 Disconnect the emission and the excitation cables and
from the detector main board.
A
B
A
B
A
EM
EX