First Steps with the Fluorescence Detector 2
1100 Series FD Reference Manual 33
Getting Started and Checkout
This chapter describes the check out of the Agilent 1100 series fluorescence
detector using the Agilent isocratic checkout sample.
Starting Your Detector
1 Turn on the detector.
2 Turn on the lamp.
When the lamp is turned on the first time the instrument performs some
internal checks and a calibration check which takes about 5 minutes.
3 You are now ready to change the settings of your detector.
When required If you want to checkout the detector
Hardware required LC system with G1321A FLD
Parts required Start-up Kit 5063-6528, includes
Column Hypersil ODS, 125 mm x 4 mm i.d. with 5 µm particles (Part number
7982618-564)
Cartridge Holder for column
Agilent isocratic checkout sample (Part number 01080-68704)
Fittings, Qty=2 (Part number 0100-1516)
Capillary, 150 mm long, 0.17 mm i.d. (Part number 5021-1817)