274 1100 Series FD Reference Manual
8 Theory of Operation
Firmware Description
The firmware of the instrument consists of two independent sections:
• a non-instrument specific section, called resident system,
• an instrument specific section, called main system.
Resident System
This resident section of the firmware is identical for all Agilent 1100 series
modules. Its properties are:
• the complete communication capabilities (GPIB, CAN, LAN and RS-232C),
• memory management,
• ability to update the firmware of the ‘main system’.
Main System
Its properties are:
• the complete communication capabilities (GPIB, CAN, LAN and RS-232C),
• memory management,
• ability to update the firmware of the resident system.
In addition the main system comprises the instrument functions that are
divided into common functions like
• run synchronization through APG remote
• error handling,
• diagnostic functions,
• or module specific functions like:
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance (see “Raw Data
Conversion to Fluorescence (LU) Units" on page 276.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals